The oracue Starter Kit includes:
- 1x oracue handpiece
- 1x 3mm guide
- 1x 2mm guide
- 1x Plate filter
- 1x 500-pack PolyAll sleeves
- 1x oracue storage case
Standard one-year limited warranty on oracue handpieces.
Cleaning & Disinfection: This product must be cleaned as part of routine preparation before each patient. Disposable sleeves are for single use only. After use, remove the sleeve. Clean the optical light guide shafts and handpiece with a wipe and 70-90% isopropyl alcohol. The body of the device may be disinfected with an intermediate-level disinfectant. Follow all disinfectant manufacturer instructions. DO NOT use bleach solutions, DO NOT immerse, DO NOT use abrasive cleaners or brushes, and DO NOT sterilize/autoclave. The Endo Guide may be disinfected as outlined above but cannot be sterilized. The Endo Fiber is for single use only and should be disposed of after each patient.